Charged Aerosol Detectors
Corona Veo and Corona Veo RS
Page 24 Operating Instructions
Connect the detector to an electrical line that shares a common ground with other system
modules. If necessary, use a properly functioning multiple socket. A common ground for
all system modules will avoid ground loops which can create erratic results (for example,
high baseline noise).
Warning: Do not use defective multiple sockets or extension cords, as
they may cause personal injury or damage to the device.
Avertissement: N'utilisez pas des défectueux blocs multiprise ou des câbles
prolongateurs. Cela pourrait entraîner des blessures
corporelles ou endommager l'instrument.
Ventilation Requirements
Ventilation must be at atmospheric pressure with no vacuum or negative pressure
applied.
Important: A vacuum in the ventilation may cause depressurization
inside the detector. This may impair detector operation and
lead to decreased performance.
Important: Un vide dans la ventilation peut conduire à la
dépressurisation à l'intérieur du détecteur. Ceci peut nuire au
bon fonctionnement du détecteur et contribuer à une
diminution des performances.
Install the HPLC system in a well-ventilated laboratory. Exhaust gases (including carrier
gas, vaporized eluents and solute micro particles) exit on the rear panel of the detector
through an external vent. The exhaust gas may contain volatile organic compounds in
low concentration. Connect the gas exhaust to a fume hood or other ventilation device
and the detector is properly vented. Observe the safety precautions on hazardous gases in
section 1.2.2 ( page 4).